Below will be the list of general procedures along with the equipment required for each testing standard we will attempt to do for testing the front each bulkhead.
Please note that although each standard says to test 5 specimens each, we don’t exactly may not have 15 specimens for the front 10 specimens of each bulkhead so we will may have to do 3/4 instead of 5less.
ASTM C393
Test Specimen Requirements
Length x Width x Depth - 200 mm x 75 mm x Thickness of Sandwich Construction
Make sure to test at least five specimens per test condition.
Equipment Needed:
Micrometers and/or callipers
Deflectometer/Deflection gages (ONLY if the UTM can’t measure the displacement)
Conditioning Chamber
Environmental Testing Chamber
3-point bend fixture
UTM
...
Variable | Quantity |
---|---|
Load rate | 6mm/min |
Failure occurence | 3-6 minutes |
Support bar separation |
ASTM D3039
Test Specimen Requirements
...
Test at least 5 specimens
Equipment Needed:
3 Force Indicator
Grips
Strain Indicating Device
Extensometer
Strain Gauge
Tabs( We should probably fabricate our own tabs)
UTM
...
Variable | Quantity |
Failure occurence | 1 to 10 min |
Head displacement | 2mm/min |
ASTM D3410
Equipment Required:
Compression fixture
Strain-indicating device
Special alignment jig
Bonded Resistance Strain Gauges
Emery cloth
I believe UTM
Conditioning chamber
Environmental test chamber
Extensometer
Procedures:
Condition specimens
Apply strain gages/extensometers to both faces of the specimen(if we choose to).
Monitor test temperature using thermocouples (if we choose to).
If needed, move the testing machine crosshead to open the distance between the 2 housing blocks so both the upper and lower wedge rip assemblies may be accessed.
Close the upper grips manually to check the specimen's vertical displacement. Do it with the lower grips as well. The upper grips should be in full contact with the wedge when closed. Repeat if we need to.
Attach strain transducers if they will be used. Then, attach the strain recording instrumentation to the strain gages or other transducers on the specimen. Then, remove any remaining preload and zero the transducer(s).
Apply the force to the fixture at the specified rate until failure while recording data.
Variable | Quantity |
---|---|
Strain Rate | 0.01 min-1 |
Crosshead Displacement | 1.5 mm/min |
Failure occurence | 1-10 min |